GE/Fuji Image
Plates at D1
Detlef
Smilgies
Mount and beamstop
- insert a plate, white or blue side facing the beam in the holder with
the photocarton cover facing upstream
- gently tighten screws, two diagonal screws are enough, top rear and bottom front screw are the best combination
- for the WAXS range the image plate should be about 200 mm from the sample, with beamstop in
the center
- use 1/16"Al as attenuator for the direct beam at 10keV; verify
beam position with a burn and without plate
- for line-up use the Idet ion chamber with beam stop about 500 mmm
downstream of sample and slide the holder out with a clamp marking the
scattering position; often Idet is already set up with the SAXS camera for GIS/WAXS
- make sure the Uniblitz shutter is closed, before opening stops to
expose a plate (shclose command)
- use macro "expose [time]" for taking images of an exposure given by [time];
monitor intensities will be saved in the scan file
- start with a 1 sec exposure for a new sample
- put exposed plate in the holder and read them (see below)
- the plate goes with the white or blue side facing up onto the holder; the magnetic backing will make it stick
- place the straight corner in A8, where the finger hole is - then plate orientation is consistent
- line up the plate with the straight edges of the holder - be careful not to bend the plate
- put the holder into the reader with the plate facing down, arrow to arrow, and close the lid
- with a bit of practice this can be done in less than 1 min
- erase plate after read-out for 5 min in the light bath
Reading the plate
- start program "Typhoon FLA 7000"
- in main menu choose mode "Phosphorimaging"
- start-up parameters: LOAD CONDITION > CHESS
- in case reader parameters are lost
- laser and correction will automatically be selected
- photo multiplier voltage (PMT): 800
- Pixel Size: 100
- Latitude: 4
- Mode (image plate size in [cm]): 20 x 25
- files for the time being are saves as TIFF and GEL - both can be read by fit2d
- image size will be about 10MB either way
- there are some remaining issues with autoscaling of saved data - to be figured out
- the Edit > Preferences menu defines the saving mode:
- Scan Settings: correction - manual, ND Filter - off, Scan Mode - Standard
- Image file settings: File Format - GEL and TIFF
- check user directory
- the H: drive is mapped to the station computer
- go to your user directory, and make a subdirectory for the images, e.g. "IP"
- choose a name of the image based on sample name and current scan number
- press "Start Scan" - the scanning menu will pop up
- be patient and wait for the reader
- some pixel statistics is provided (inset on the right) - you can judge from that how close you are to saturation
- use "linear range" for viewing
- the left and right red lines indicating the grayscale range
can be moved with the mouse to change the image contrast and will show
up in the grayscale image
- the image on the left hand provides a gray scale image of the exposure
- some additional remarks
- you can "Stop" reading the plate if necessary (wrong plate etc.)
- you do not need to use "Save" - the original image is already saved
- you do not need to use "Launch" - the program launched is for fluorescent assays
- wait until the "Return" botton is active; now the plate can be safely removed
- images can be analyzed with fit2d (version 12)
- image size is 2500 x 2000 pixels
- typical range (user min/max) for TIFF is 65 to 65,000
- GEL is a horizontally mirrored image of TIFF; have not figured out range
- image autoscaling needs to be reviewed - not enough documentation available