Saturation
Effects in Linear
Gasdetectors
Detlef
Smilgies, CHESS
Detection
Principle
A linear gasdetector is a
position-sensitive proportional counter. A
strong electric field between counting wire and anode plate (2000V over
5-10 mm) leads to a charge avalanche when an x-ray photon ionizes an
atom of the counter gas (typically Ar or Xe at 1 to 5 bar).
Because of a quenching gas (often 10% methan; the mix with Ar is called
P10) the discharge dies after a brief time and after creating ample
charge ("charge amplification"). The charge pulse is detected at either
end of the wire. Either the charge ratio (resistive current division -
MBraun detector) or the delay time between the arrival times of both
pulses is used to determine the position of the primary ionization
event along the wire.
Saturation
Effects
-
integral effect
The saturation of the integral
counts along the wire is often a
property of the detection electronics. While the time resolution of
charge dividing detection is given by the proper integration times of a
few microsec, typical delay lines have time constants around several
100 ns. The randomness of the Poisson distributed counts will limit
linear detection rates to 1/(100*time constant). G2's
Time-to-Amplitude-Converter - Pulse-Height-Analyzer
combination (TAC-PHA) will saturate at about 10 kcps, mostly given by
the limited
speed of the ORTEC PCI board. A good Time-to-Digital Converter (TDC)
can go up to 100 kcps. The
dead time correction follows the usual formula, and spectra can be
corrected.
-
local effect
If the detector sweeps over a
strong peak, so much space charge can
build up locally that the peak saturates or even turns into an inverted
peak. On the scope this behavior is associated in a reduction of the
peak amplitude from the sum output (the energy-proportional signal).
The means, that the gas amplification becomes reduced which causes
problems with threshold values in the electronics. Moreover the overall
signal rate arounf the peak seems to be affected. I am not aware
of any correction procedure other than remeasuring a range of spectra
with a suitable attenuator. The overall signal along the wire may also
be affected, so spectra with local saturation can not be trusted much.
By carefully tuning the gas amplification the detector can be optimized
using low-noise preamps.
-
wire degradation
If the detector is hit hard at the
same spot over and over again, the
detection sensitivity at this spot may be strongly reduced. This is due
to carbon deposits from cracking methane in the discharge. This problen
often occurs at the direct beam position used for line-up. Hence care
must be used to line-up in the direct beam with a sufficint attenuation
yielding integral count rates below 1kcps. Only remedy is to renew the
counting wire.
Conclusion
Gas detectors are elegant,
relatively simple, and cost efficient x-ray
detectors, as long as they are operated in a good working range
specified by a maximum integral and a maximum peak countrate.
Unfortunately,
these numbers are often not specified by the manufacturer, and it
usually
means some painful, hard-learning exercise for the experimenter.
Moreover, the experimenter usually deals with a "black box", and it is
not very obvious how to adjust detector parameters such as the gas
amplification, in order to set up a convenient working range of the
detector.
For G2 we recently ordered a
linear diode array from the BNL detector
group which can go up to 100 kcps per diode element. This monstrous
performance is achieved in a custom high-integrated circuit providing a
preamp and discriminator for each of the 1000 elements. The new
detector, expected to materialize sometime in 2008, will be much better
adapted to the flux and dynamic range of signals at G2.