MedOptics CCD @ G3 - Check List ================================= I line-up A close main shutter B ccdoff C mv del 0 gam 0 D move CCD back E cover CCD F remove lead mask on Be window G mount ion chamber and slit H horizontal slit gap should be 0.5 mm, vertical gap 2 mm I shopen J line up in the direct beam a mv mu 0 zeta 0 chamy -1 att 3 b dscan mu -1 1 20 1; mv mu CEN; set mu 0 c dscan del -1 1 20 1; mv del CEN; set del 0 d mv chamy 0; dscan chamy -1 1 20 1; mv chamy CEN; set chamy 0 e dscan zeta -1 1 20 1; mv zeta CEN; set zeta 0 K line up incident angle with the reflected beam a mv mu 0.3 zeta -0.15 b dscan zeta -.2 .2 20 1; mv zeta CEN; set zeta 0.15; laser c dscan chamy -1 1 20 1; mv chamy PEAK; laser d put print-outs of final line-up scans in the log II CCD measurements A ccdon B remove ion chamber and slit C tape lead mask onto Be window using the marks on the window edge D remove CCD cover E move CCD close to Be window using marks on the dove tail F make sure all six plastic screws of the CCD holder are hand-tight G mv del 18.5 gam 2 att 0 H take some typical dark images: dark 0.1;dark 1;dark 5;dark 10 I take a test images to check set-up: take 0.1; take 1; take 10 J take images to record scattering background: take 1;take 5;take 10 K start growth: tseries 100 10 L monitor CCD intensity and reduce exposure if needed III post-mortem CCD scans A chamber y scan: dscan chamy -.4 .4 8 10 move chamy to best position B incident angle scan: mv del 18.5; dscan zeta -0.04 0.04 8 10 move zeta to best position C GID series of grown film: ascan del 13.5 38.5 5 10 D angle calibration: mv del 17.5 gam 2;take 10 mv del 18.5 gam 1;take 10 mv del 18.5 gam 2 E check deposited film size: ascan samz -8 7 15 10