RESETTING TRIPS IN MUONS

1. Open the  Detector-Based Adjustment Frame GUI
2. If there are channels listed in the lower half of the window, press �Clear All� for 
   the system they belong to.
3. Select MUON Inner, Outer, Endcap, and Return.
4. Right click on BEN column and change value to 0 to disable bulk.
5. Hit �Refresh� several times while you wait for BMV column to go below 100 Volts on 
   all chambers.
6. Go to Muon HV Bleeder window.
7. Type BLEED at the prompt.
8. Wait for 5 seconds.
9. Type CLOT to remove Muons from bleeder circuit.
10. Return to Detector-Based Adjustment Frame GUI.
11. Right click on CEN column and change value to 1 to reset the trip.
12. Hit refresh several times and make sure trip is reset.  
13. Right click on BEN column and change value to 1 to ramp up the MUON HV.