04/10/2007

Remove the West Slot 7 board

During the access time this morning, I took the HV board(HV#37) on Pit West Slot 7. Then we run the HV server with Tag : HVTest_noWS7 without seeing any problem. After that, I jumped the patch panel in the West to power Layer 45 by 34, 46 by 35, and 47 by 36. i.e. S7 by S6 in the West.

We tested the server by rumping to FILL and RUN and didn't see any problem.

One thing need to mention is that the sector in Layer 34, channels 16-18 are now powering up four sectors in total. More precisely, it powers L34-ch16,17,18, L34-ch19,20,21, L45-ch16,17,18, L45-ch19,20,21. By far, no extra currents noticed on those channels.

Update the HV Module Board log: (HV37)

04/11/2007

Update HVNormal

Cannot open the SessionManager on solon2.

Finish the work at the counting room.

04/20/2007

HV Board 37

Taken back from Margie.

04/24/2007

Replace #45 board with #37 from HVEast slot 5

Due to the occasional non-recognition problem happened on HVEast slot 5, I took out that board (#45) and replaced it with the good spare board #37. I have sent the #45 board to electric shop for cleaning and will investigate it after taking back.

All of the board status have been updated on page:

http://www.lepp.cornell.edu/~xs32/private/hv/ModuleLog.html

Wrote to e-log.

04/30/2007

Update spare boards info

done.

Prepare tag for sending out board 16,17,46

-------------------------------------------------------------------
HVBoard #16
-------------------------------------------------------------------

Symptom:

 - Bulk 0,1 ramp to 2.5kV overload

 - Bulk 2 trips when ramping (with load or w/out).
   If bulk 1 is under load and bulk 2 enabled, it trips at ~2300V.



-------------------------------------------------------------------
HVBoard #17
-------------------------------------------------------------------

Symptom:

 - Not recognized by mainframe on boot-up.



-------------------------------------------------------------------
HVBoard #46
-------------------------------------------------------------------

Symptom:

 - Mainframe does not recorgnize module.

 - Pulls down voltage on backplane. 800 kHz clock to half ampl.
   SI signal disappears.


-------------------------------------------------------------------

Print out and place them on board and envelope.